REAL-SAMPLE HURDLES

What changes outside the buffer

๐Ÿฉธ

Complex matrix

Blood, saliva, or food add proteins, salts, and particles

๐Ÿง‚

High salt

Physiological salt shortens the Debye length

๐Ÿงด

Fouling

Non-specific binding builds up over minutes

Every earlier limit gets harder in a real sample.

SAMPLE PREP

Making the sample easier

Filter Removes cells and particles
Dilute Lowers salt and background proteins together
Desalt Swaps the buffer for a low-salt one before measurement
Concentrate Raises a scarce target's concentration before it reaches the sensor
Prep steps trade convenience for a cleaner signal.

RELIABILITY

Sensors must agree with each other

device-to-device variation

Two nominally identical BioFETs rarely respond identically. Threshold voltages, receptor densities, and noise all vary. Reliable results need calibration per device or per batch, replicates, and reference sensors on the same chip.

Calibrate, replicate, and always run a reference.

POINT OF CARE

What users need

๐Ÿ’ฐ

Affordable

A low cost per test

โšก

Rapid and robust

A fast answer that tolerates rough handling

๐ŸŽ›๏ธ

User-friendly

Little or no training required

๐Ÿ“ฆ

Deliverable

Usable where the patient is, without lab infrastructure

These are four of the seven ASSURED criteria for point-of-care tests.

VALIDATION

A checklist before claiming a result

Calibration curve Known concentrations, fitted, with error bars
Selectivity Blanks and wrong targets give no response
Reproducibility Several devices and batches agree
Real sample The same response in a spiked real matrix, not just in buffer
A result is only as strong as its weakest control.

UNIT 25 STUDY COMPLETE

๐Ÿ”ฌ

Ready for the Fab Challenge?

You've covered what it takes to turn a working prototype into a dependable measurement.